The 50 Turn 5500A/COIL is used as a tool for calibrating clamp type current meters that operate by two different principles ? as current transformers (AC only), and by the Hall Effect (Both AC and DC). It is impractical to calibrate 1000A rated current clamp meters using a 1000A source. However, by using the 50 turns of the 5500A/COIL in conjunction with a current calibrator you can effectively multiply the current of the current calibrator by a factor of 50 to support the calibration and verification of these clamp-on current meters.
型號名稱 描述
5500A/COIL 50 匝電流線圈
本公司長期高價收購/回收:Agilent、HP、Anritsu、Advantest、R/S、/MARCONI網(wǎng)絡分析儀、手機綜合測試儀、無線電綜合測試儀、萬用表校準儀、頻譜分析儀、視頻分析儀、音頻分析儀、噪聲系數(shù)分析儀、藍牙測試儀、視頻圖象信號源、LCR測試儀(電橋)、高低頻信號源、示波器、萬用表、手機程控電源、電子負載、頻率計、功率計、儀器配件,GPIB卡、GPIB轉USB線等等。歡迎來電咨詢!!!
Supports the calibration of clamp-type current meters
The 50 Turn 5500A/COIL is used as a tool for calibrating clamp type current meters that operate by two different principles ? as current transformers (AC only), and by the Hall Effect (Both AC and DC). It is impractical to calibrate 1000A rated current clamp meters using a 1000A source. However, by using the 50 turns of the 5500A/COIL in conjunction with a current calibrator you can effectively multiply the current of the current calibrator by a factor of 50 to support the calibration and verification of these clamp-on current meters
General Specifications
Number of Turns
50
Maximum Current
11A rms, continuous
20A rms, 2 minutes
Maximum Duty Cycle Derating
< 11A, continuous
> 11A, 2 minutes ON, 8 minutes OFF
Maximum Voltage
3V rms
Frequency of Operation
DC, 45-440 Hz
(should not exceed rms Voltage rating)
For the 5500A Current Output:
DC, 45-65 Hz, 0-11A
65-440 Hz, 0-2.19999A
For the 5520A Current Output:
DC, 45-65 Hz, 0-20A
65-440 Hz, 0-2.99999A
This specification assumes that a Fluke 80i-600 Clamp Meter or equivalent is attached. Other clamp meters may limit the 5500A/5520A Current Output drive capability to less than the above.
Uncertainty Due to Clamp Meter/Coil Interaction
± (0.25% of effective output + 0.5A), for toroidal-wound current clamps, such as the Fluke 80I and 80I-1000.
± (0.50% of effective output + 0.5A), for current clamps like the Fluke 80i-kw, 80i-400, 80i-410, 80i-500, 80i-1010, Fluke 31, Fluke 33, or equivalent.
Minimum Inside Diameter of Clamp Jaws
2.54 cm (1 in.)